Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2018
Comparison of two different radiofrequency ablation systems for renal artery denervation: Evaluation of short-term and long-term follow up.
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Al Raisi Sara I, Pouliopoulos Jim, Qian Pierre, King Patricia, Byth Karen, Barry Michael T, Swinnen John, Thiagalingam Aravinda, Kovoor Pramesh
Аннотация
OBJECTIVES: To assess the clinical efficacy of renal artery denervation (RAD) in our center and to compare the efficacy of two different radiofrequency (RF) systems.
BACKGROUND: Several systems are available for RF renal denervation. Whether there is a difference in clinical efficacy among various systems remains unknown.
METHODS: Renal artery denervation was performed on 43 patients with resistant hypertension using either the single electrode Symplicity Flex (n = 20) or the multi-electrode EnligHTN system (n = 23). Median post-procedural follow-up was 32.93 months. The primary outcome was post-procedural change in office blood pressure (BP) within 1 year (short-term follow-up). Secondary outcomes were change in office BP between 1 and 4 years (long-term follow-up) and the difference in office BP reduction between the two systems at each follow-up period.
RESULTS: For the total cohort, mean baseline office BP (systolic/diastolic) was 174/94 mmHg. At follow-up, mean changes in office BP from baseline were -19.70/-11.86 mmHg (P < 0.001) and -21.90/-13.94 mmHg (P < 0.001) for short-term and long-term follow-up, respectively. The differences in office BP reduction between Symplicity and EnligHTN groups were 8.96/1.23 mmHg (P = 0.42 for systolic BP, P = 0.83 for diastolic BP) and 9.56/7.68 mmHg (P = 0.14 for systolic BP, P = 0.07 for diastolic BP) for short-term and long-term follow-up, respectively.
CONCLUSIONS: In our cohort, there was a clinically significant office BP reduction after RAD, which persisted up to 4 years. No significant difference in office BP reduction between the two systems was found.
BACKGROUND: Several systems are available for RF renal denervation. Whether there is a difference in clinical efficacy among various systems remains unknown.
METHODS: Renal artery denervation was performed on 43 patients with resistant hypertension using either the single electrode Symplicity Flex (n = 20) or the multi-electrode EnligHTN system (n = 23). Median post-procedural follow-up was 32.93 months. The primary outcome was post-procedural change in office blood pressure (BP) within 1 year (short-term follow-up). Secondary outcomes were change in office BP between 1 and 4 years (long-term follow-up) and the difference in office BP reduction between the two systems at each follow-up period.
RESULTS: For the total cohort, mean baseline office BP (systolic/diastolic) was 174/94 mmHg. At follow-up, mean changes in office BP from baseline were -19.70/-11.86 mmHg (P < 0.001) and -21.90/-13.94 mmHg (P < 0.001) for short-term and long-term follow-up, respectively. The differences in office BP reduction between Symplicity and EnligHTN groups were 8.96/1.23 mmHg (P = 0.42 for systolic BP, P = 0.83 for diastolic BP) and 9.56/7.68 mmHg (P = 0.14 for systolic BP, P = 0.07 for diastolic BP) for short-term and long-term follow-up, respectively.
CONCLUSIONS: In our cohort, there was a clinically significant office BP reduction after RAD, which persisted up to 4 years. No significant difference in office BP reduction between the two systems was found.