Caution needed in using oral polio vaccine beyond the cold chain: vaccine vial monitors may be unreliable at high temperatures.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Shrivastava Ashutosh, Gupta Neeraj, Upadhyay Pramod, Puliyel Jacob
Аннотация
BACKGROUND & OBJECTIVES: Stabilized live attenuated oral polio vaccine (OPV) is used to immunize children up to the age of five years to prevent poliomyelitis. It is strongly advised that the cold-chain should be maintained until the vaccine is administered. It is assumed, that vaccine vial monitors (VVMs) are reliable at all temperatures. VVMs are tested at 37°C and it is assumed that the labels reach discard point before vaccine potency drops to >0.6 log10. This study was undertaken to see if VVMs were reliable when exposed to high temperatures as can occur in field conditions in India.
METHODS: Vaccine vials with VVMs were incubated (10 vials for each temperature) in an incubator at different temperatures at 37, 41, 45 and 49.5°C. Time-lapse photographs of the VVMs on vials were taken hourly to look for their discard-point.
RESULTS: At 37 and 41°C the VVMs worked well. At 45°C, vaccine potency is known to drop to the discard level within 14 h whereas the VVM discard point was reached at 16 h. At 49.5°C the VVMs reached discard point at 9 h when these should have reached it at 3 h.
CONCLUSION: Absolute reliance cannot be placed on VVM in situation where environmental temperatures are high. Caution is needed when using 'outside the cold chain' (OCC) protocols.
METHODS: Vaccine vials with VVMs were incubated (10 vials for each temperature) in an incubator at different temperatures at 37, 41, 45 and 49.5°C. Time-lapse photographs of the VVMs on vials were taken hourly to look for their discard-point.
RESULTS: At 37 and 41°C the VVMs worked well. At 45°C, vaccine potency is known to drop to the discard level within 14 h whereas the VVM discard point was reached at 16 h. At 49.5°C the VVMs reached discard point at 9 h when these should have reached it at 3 h.
CONCLUSION: Absolute reliance cannot be placed on VVM in situation where environmental temperatures are high. Caution is needed when using 'outside the cold chain' (OCC) protocols.