Does the amplatzer septal occluder device alter ventricular contraction pattern? A ventricular motion analysis by MR tagging.

Stern Heiko, Baurecht Hansjörg, Luechinger Roger, Schwaiger Markus, Hess John, Fratz Sohrab

Контролируемое клиническое исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: To assess by cardiovascular magnetic resonance (CMR) and CMR tagging if the Amplatzer Septal Occluder affects right ventricular (RV) and left ventricular (LV) motion pattern.

MATERIALS AND

METHODS

Sixteen consecutive patients with significant atrial septal defect (ASD) and nine consecutive patients with persistent foramen ovale (PFO) as controls were studied before and a median of 14 days after defect closure by an Amplatzer occluder. By CMR end-diastolic (EDV) and end-systolic (ESV) RV and LV volumes were determined. Aortic and pulmonary artery flow was measured for assessment of left-to-right shunt (Qp/Qs). By CMR tagging circumferential strain and radial shortening, maximal rotation and torsion were measured,

RESULTS

In ASD patients RV-EDV and RV-ESV decreased (P < 0.05). LV-EDV and LV-ESV increased after ASD closure (P < 0.005). Qp/Qs dropped from 1.8 to 1.0 (P < 0.001). PFO patients showed no ventricular volume change after PFO closure. In ASD patients circumferential strain and radial shortening and maximal rotation of the RV decreased by ASD closure (P < 0.01). In LV only maximal rotation at the base and apex decreased significantly (P < 0.05). Torsion remained constant. In PFO patients no tagging parameter changed after defect closure.

CONCLUSION

The Amplatzer occluder itself does not change the ventricular contraction pattern. All volume and myocardial deformation changes were caused by ventricular loading shifts.

Для цитирования: Stern Heiko, Baurecht Hansjörg, Luechinger Roger, Schwaiger Markus, Hess John, Fratz Sohrab. Does the amplatzer septal occluder device alter ventricular contraction pattern? A ventricular motion analysis by MR tagging.. Journal of magnetic resonance imaging : JMRI. 2012. DOI: 10.1002/jmri.23575.