BACKGROUND

Automatic continuous positive airway pressure (automatic CPAP, APAP) is an effective treatment option in the obstructive sleep apnoea syndrome (OSAS). The differentiation of obstructive and central respiratory events is crucial in adjusting the optimal pressure in this treatment mode. In this pilot study we evaluated a new automatic CPAP algorithm in OSAS patients.

METHODS

14 patients with newly diagnosed obstructive sleep apnoea syndrome were enrolled. After a diagnostic polysomnography, patients were treated for one night with a new APAP device based on flow, snoring, relative minute volume and the obstructive pressure peak signal.

RESULTS

The total apnoea/hypopnoea index (AHI) was 30.0 +/- 21.4/h at baseline and 3.7 +/- 5.3/h with APAP ( P < 0.005). Both obstructive AHI (22.7 +/- 20.5/h at baseline, 1.5 +/- 3.5/h with APAP, P < 0.005) and central AHI (7.3 +/- 4.9/h and 2.2 +/- 2.5/h, respectively, P < 0.01) as well as the arousal index (25.4 +/- 18.1/h and 5.1 +/- 3.8/h, respectively, P < 0.005) were reduced significantly with the new algorithm.

CONCLUSIONS

The new algorithm of an automatic CPAP device is effective in the treatment of obstructive sleep apnoea syndrome.

Для цитирования: Galetke W, Stieglitz S, Anduleit N, Kenter M, Kühnel J, Osagie-Paech R, Richter K, Randerath W. [Evaluation of a new automatic CPAP algorithm in the treatment of obstructive sleep apnoea syndrome].. Pneumologie (Stuttgart, Germany). 2009. DOI: 10.1055/s-0028-1119519.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.