MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/jmri.23520

Медицинская библиотека · Journal of magnetic resonance imaging : JMRI · 2011

A deflectable guiding catheter for real-time MRI-guided interventions.

Bell Jamie A, Saikus Christina E, Ratnayaka Kanishka, Wu Vincent, Sonmez Merdim, Faranesh Anthony Z, Colyer Jessica H, Lederman Robert J, Kocaturk Ozgur

СтатьяResearch Support, N.I.H., IntramuralПеревод в очереди

Аннотация

PURPOSE: To design a deflectable guiding catheter that omits long metallic components yet preserves mechanical properties to facilitate therapeutic interventional MRI procedures.

MATERIALS AND

METHODS

The catheter shaft incorporated Kevlar braiding. A 180° deflection was attained with a 5-cm nitinol slotted tube, a nitinol spring, and a Kevlar pull string. We tested three designs: passive, passive incorporating an inductively coupled coil, and active receiver. We characterized mechanical properties, MRI properties, RF induced heating, and in vivo performance in swine.

RESULTS

Torque and tip deflection force were satisfactory. Representative procedures included hepatic and azygos vein access, laser cardiac septostomy, and atrial septal defect crossing. Visualization was best in the active configuration, delineating profile and tip orientation. The passive configuration could be used in tandem with an active guidewire to overcome its limited conspicuity. There was no RF-induced heating in all configurations under expected use conditions in vitro and in vivo.

CONCLUSION

Kevlar and short nitinol component substitutions preserved mechanical properties. The active design offered the best visibility and usability but reintroduced metal conductors. We describe versatile deflectable guiding catheters with a 0.057" lumen for interventional MRI catheterization. Implementations are feasible using active, inductive, and passive visualization strategies to suit application requirements.

Для цитирования: Bell Jamie A, Saikus Christina E, Ratnayaka Kanishka, Wu Vincent, Sonmez Merdim, Faranesh Anthony Z, Colyer Jessica H, Lederman Robert J, Kocaturk Ozgur. A deflectable guiding catheter for real-time MRI-guided interventions.. Journal of magnetic resonance imaging : JMRI. 2011. DOI: 10.1002/jmri.23520.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.