MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jhin.2018.03.001

Медицинская библиотека · The Journal of hospital infection · 2018

Reducing the risk of iatrogenic Creutzfeldt-Jakob disease by improving the cleaning of neurosurgical instruments.

Smith A, Winter S, Lappin D, Sherriff A, McIvor I, Philp P, Suttner N, Holmes S, Stewart A

СтатьяПеревод в очереди

Аннотация

BACKGROUND

In all, there have been 178 variant Creutzfeldt-Jakob disease (vCJD) patients diagnosed in the UK, with an estimated maximum 1:2000 carriage rate based on archived appendix and tonsil tissue, implying that infection may be rare but carriage relatively frequent. Previous workers have identified that maintenance of surgical instruments in a humid atmosphere after use and prior to cleaning assists cleaning efficacy. Recently the Department of Health/Advisory Committee on Dangerous Pathogens UK have recommended a surgical instrument cleanliness threshold post cleaning of <5 μg protein per instrument side.

AIM: To quantify cleanliness of neurosurgical instruments and to investigate cost-effective measures for improved cleaning.

METHODS

Two instrument protein quantification methods were used: one based on the International Standard (15883 series) using sodium dodecyl sulphate elution and ortho-phthalaldehyde reaction, and a second in-situ protein fluorescence detection system (ProReveal) providing results per instrument side. In-vitro investigation of the efficacy of some commercial and in-house pre-clean wetting agents was undertaken using artificial test soil and stainless steel discs under standard conditions. In-vivo evaluation of best-performing in-vitro agents was undertaken on craniotomy sets.

FINDINGS: ProReveal technology demonstrated that 163 out of 187 (87%) neurosurgical instruments had <5 μg residual protein per instrument side. The use of proprietary National Health Service plastic bags and sterile water-soaked wound pads were equivalent in efficacy to commercial pre-cleaning wetting products and significantly less expensive.

CONCLUSION

Although we demonstrate low in-situ protein levels on neurosurgical instruments and the beneficial effects of keeping instruments moist, other cleaning critical-control points such as instrument loading patterns should also be monitored.

Для цитирования: Smith A, Winter S, Lappin D, Sherriff A, McIvor I, Philp P, Suttner N, Holmes S, Stewart A. Reducing the risk of iatrogenic Creutzfeldt-Jakob disease by improving the cleaning of neurosurgical instruments.. The Journal of hospital infection. 2018. DOI: 10.1016/j.jhin.2018.03.001.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.