European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery · 2021
The efficacy of sinus plication in aortic valvuloplasty for bicuspid aortic valve: experiments in a pulsatile flow simulation model.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Arimura Satoshi, Takada Jumpei, Nishimura Gohki, Nakama Natsuki, Kawasaki Eita, Matsuhama Minoru, Iwasaki Kiyotaka, Kasegawa Hitoshi, Kunihara Takashi
Аннотация
OBJECTIVES: Sinus plication has emerged as a promising tool that can lead to better stability in bicuspid aortic valve (BAV) repair. However, the mechanisms underlying the efficacy of this technique are unclear. We evaluated the hydrodynamic effect of sinus plication using the experimental pulsatile flow simulator and our original BAV model in vitro.
METHODS: Based on the computed tomography data of a BAV patient who had undergone aortic valvuloplasty, a BAV model (group C, n = 6) was developed with bovine pericardium and vascular prosthesis (J-graft Shield Neo Valsalva 24 mm). We performed sinus plication (group SP, n = 6) in the BAV model and compared hydrodynamic data with the control model in the pulsatile flow simulator. Non-fused cusp angle, annulus diameter and effective height were measured by ultrasonography.
RESULTS: The average flow was significantly increased in group SP compared to group C (4.24 ± 0.14 l/min vs 4.14 ± 0.15 l/min, respectively, P = 0.034). The mean transvalvular pressure gradient and regurgitant fraction were significantly decreased in group SP compared to group C (11.6 ± 4.3 mmHg vs 16.6 ± 5.0 mmHg, respectively, P = 0.009 and 14.1 ± 2.0% vs 17.4 ± 2.1%, respectively, P = 0.001). Ultrasound measurement indicated that non-fused cusp angle was significantly increased in group SP compared to group C (163.8° ± 9.2° vs 153.0° ± 4.6°, respectively, P = 0.012).
CONCLUSIONS: Sinus plication in the BAV model significantly increased the commissural angle. It was effective in not only controlling regurgitation but also improving valve opening. These finding should be confirmed by evaluating cusp stress and/or long-term durability in the future studies.
METHODS: Based on the computed tomography data of a BAV patient who had undergone aortic valvuloplasty, a BAV model (group C, n = 6) was developed with bovine pericardium and vascular prosthesis (J-graft Shield Neo Valsalva 24 mm). We performed sinus plication (group SP, n = 6) in the BAV model and compared hydrodynamic data with the control model in the pulsatile flow simulator. Non-fused cusp angle, annulus diameter and effective height were measured by ultrasonography.
RESULTS: The average flow was significantly increased in group SP compared to group C (4.24 ± 0.14 l/min vs 4.14 ± 0.15 l/min, respectively, P = 0.034). The mean transvalvular pressure gradient and regurgitant fraction were significantly decreased in group SP compared to group C (11.6 ± 4.3 mmHg vs 16.6 ± 5.0 mmHg, respectively, P = 0.009 and 14.1 ± 2.0% vs 17.4 ± 2.1%, respectively, P = 0.001). Ultrasound measurement indicated that non-fused cusp angle was significantly increased in group SP compared to group C (163.8° ± 9.2° vs 153.0° ± 4.6°, respectively, P = 0.012).
CONCLUSIONS: Sinus plication in the BAV model significantly increased the commissural angle. It was effective in not only controlling regurgitation but also improving valve opening. These finding should be confirmed by evaluating cusp stress and/or long-term durability in the future studies.