DOI 10.1097/md.0000000000026770
Ventriculoperitoneal shunt malfunction diagnosis based on substance dilution.
Аннотация
OBJECTIVE
Current methods for the diagnosis of ventriculoperitoneal (VP) shunt malfunction lack specific standards; therefore, it may be missed or misdiagnosed. Hence, providing a reliable diagnostic method will help improve the accuracy of preoperative decision-making. Therefore, the aim of the study was to provide a new method for the diagnosis of VP shunt malfunction.
METHODS
After in vitro testing, we enrolled a total of 12 patients with VP shunt malfunction. Before revision surgery, 0.1 mL of a 5% sodium valproate (SV) solution was injected into the reservoir; 0.1 mL of the cerebrospinal fluid (CSF) was withdrawn 20 minutes later from the reservoir to measure the SV concentration. The process was repeated on the seventh day after surgery and compared with the preoperative results.
RESULTS
The mean ± standard deviation preoperative SV concentration in the cerebrospinal fluid was greater than the postoperative concentration (5967.8 ± 1281.3 vs 391.1 ± 184.6 μg/mL, P = .001).
CONCLUSION
The proposed method is a reliable, safe, and relatively simple alternative for the diagnosis of VP shunt malfunction and further provides a reference for treatment.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.