Аннотация

BACKGROUND

Neointimal hyperplasia causes a high rate of hemodialysis synthetic graft failure. Thus, therapies that inhibit neointimal hyperplasia are urgently needed. The Coll-R is a sirolimus-eluting collagen matrix designed for intra-operative perivascular implantation around the graft-venous anastomosis. Sirolimus is an anti-proliferative drug that has proven clinical utility in suppressing neointimal tissue growth in coronary artery disease when delivered locally to the vascular wall by an endovascular drug eluting stent.

METHODS

A cohort of 12 chronic hemodialysis patients underwent surgical placement of 13 polytetrafluoroethylene grafts + Coll-R and were followed for up to 24 months. The primary endpoint was safety (freedom from device related adverse events). Secondary endpoints were pharmacokinetics of sirolimus release, success of Coll-R implantation and primary unassisted graft patency.

RESULTS

There were no technical failures, infections, vascular anastomotic or wound-healing problems. Whole blood sirolimus levels rose to a mean peak of 4.8 ng/mL at 6 h and fell to <1 ng/mL at 1 week (n = 5). Twelve and 24-month primary unassisted patencies were 76 and 38%, respectively, and the thrombosis rate was 0.37/patient-year.

CONCLUSIONS

Perivascular implantation of the Coll-R during graft surgery safely delivered sirolimus to the vascular wall. Systemic sirolimus levels were sub-therapeutic for immunosuppression. This small first-in-human study supports the concept that the Coll-R can safely deliver sirolimus to the graft-venous anastomosis. Safety and patency in this small study were sufficiently encouraging to justify randomized controlled trials to further test the efficacy of the Coll-R.

Для цитирования: Paulson William D, Kipshidze Nicholas, Kipiani Konstantine, Beridze Nutsa, DeVita Maria V, Shenoy Surendra, Iyer Sriram S. Safety and efficacy of local periadventitial delivery of sirolimus for improving hemodialysis graft patency: first human experience with a sirolimus-eluting collagen membrane (Coll-R).. Nephrology, dialysis, transplantation : official publication of the European Dialysis and Transplant Association - European Renal Association. 2012. DOI: 10.1093/ndt/gfr667.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.