MedElevator Войти в исследования
← Поиск публикаций
Медицинская библиотека · The Journal of urology · 1999

A remote controlled intraurethral insert for artificial voiding: a new concept for treating women with voiding dysfunction.

Madjar S, Sabo E, Halachmi S, Wald M, Issaq E, Moskovitz B, Beyar M, Nativ O

Клиническое исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: Many patients with voiding dysfunction find clean intermittent catheterization unsatisfactory. We evaluated the efficacy of the new remote controlled intraurethral In-Flow Catheter insert for treating women with voiding dysfunction.

MATERIALS AND

METHODS

We evaluated the efficacy of the intraurethral insert in 92 women with a mean age of 56 years. The insert is available in various sizes to adapt to individual urethras. It comprises a valve and pump assembly. A remote control unit is operated to open the valve and activate the pump, generating active urine flow.

RESULTS

In 45 patients (49%) the device was removed after a mean of 7.1 days due to local discomfort or urinary leakage around the insert. At a mean followup of 7.6 months (range 2 to 26) 47 women (51%) continued to use the device, and all are dry with complete bladder emptying. The insert was replaced periodically at a mean of 38 days to prevent salt deposits in and around it which lead to urine leakage. Asymptomatic bacteriuria developed in 22 patients (46.8%). Clinical urinary tract infections resolved in 3.9% of all patients-months with oral antibiotics. No dyspareunia was reported. All patients were satisfied with the insert and preferred it to previous treatment modalities.

CONCLUSIONS

This new remote controlled intraurethral insert is safe and effective in women with voiding difficulties.

Для цитирования: Madjar S, Sabo E, Halachmi S, Wald M, Issaq E, Moskovitz B, Beyar M, Nativ O. A remote controlled intraurethral insert for artificial voiding: a new concept for treating women with voiding dysfunction.. The Journal of urology. 1999.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.