Complicated Holter tracing with an incidence of inappropriate mode switch due to sensing abnormalities in a patient with dual-chamber pacemaker.

Poręba Małgorzata, Karczmarewicz Stefan, Szwarc Bartłomiej, Sobieszczańska Małgorzata, Wranicz Jerzy Krzysztof

Клинический случайСтатьяПеревод в очереди

Аннотация

24-hour Holter monitoring of a 59 year-old man with DDDR pacemaker (programmed mode: DDD) implanted for sick sinus syndrome with paroxysmal atrial fibrillation was performed one month after implantation, due to palpitations. Several episodes of rapid pacing of decreasing rate were detected. Intracardiac recording stored in the pacemaker memory had shown episodes of atrial lead oversensing, which led to the mode switch, resulting in DDIR mode. Signals which caused oversensing were not seen in Holter tracing.

Для цитирования: Poręba Małgorzata, Karczmarewicz Stefan, Szwarc Bartłomiej, Sobieszczańska Małgorzata, Wranicz Jerzy Krzysztof. Complicated Holter tracing with an incidence of inappropriate mode switch due to sensing abnormalities in a patient with dual-chamber pacemaker.. Cardiology journal. 2011.