m+Medelevator.RU
← Поиск публикаций
Mayo Clinic proceedings · 2004

Transcatheter Amplatzer device closure of atrial septal defect and patent foramen ovale in patients with presumed paradoxical embolism.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Khositseth Anant, Cabalka Allison K, Sweeney John P, Fortuin F David, Reeder Guy S, Connolly Heidi M, Hagler Donald J

Аннотация

OBJECTIVE: To review our experience with, and profile the safety and efficacy of, the Amplatzer PFO (patent foramen ovale) occluder (APO) and Amplatzer septal occluder (ASO) used to close PFO and/or atrial septal defect (ASD) in patients with paradoxical embolism (PE).

PATIENTS AND METHODS: Between April 1998 and November 2002, 103 patients at the Mayo Clinic in Rochester, Minn, and Scottsdale, Ariz, mean age 52.4 years, with presumed PE (transient ischemic attack [n=22], stroke [n=77], or peripheral emboli [n=4]) underwent transcatheter device closure of PFO (n=81), ASD (n=12), and ASD/PFO (n=10) with 106 devices (APO [n=22] or ASO [n=84]).

RESULTS: All devices deployed successfully, and no patients died. Procedural complications included atrial fibrillation (n=2), vessel injury (n=3), profound sinus node dysfunction (n=1), and device embolization with successful retrieval (n=1). At 3 months, 7 of 95 monitored patients had trivial residual shunt; at 12 months, 2 of 28 monitored patients had trivial residual shunt. Three patients had recurrent events--2 transient ischemic attacks and 1 retinal artery occlusion--at a mean +/- SD follow-up of 8.3 +/- 8.1 months (range, 1-34 months). None of these 3 patients had residual shunt or evidence of intracardiac thrombus. The average annual recurrence of all events was 3.6% at 23 months. The overall mean +/- SD freedom from recurrence of all events was 98.9% +/- 1.2% and 83.8% +/- 10.2% at 12 and 29 months of follow-up, respectively.

CONCLUSIONS: Transcatheter device closure of PFO and/or ASD with use of APO/ASO in patients with presumed PE is effective and safe. Recurrent events may occur in the absence of a residual shunt.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты