Accuracy of the OMRON HEM-7361T blood pressure monitor in the differentiation between atrial fibrillation and sinus rhythm.

Zhou Yi, Xia Jia-Hui, Zhang Wei, Li Yan, Wang Ji-Guang

СтатьяКлиническое исследованиеПеревод в очереди

Аннотация

OBJECTIVE

We investigated the accuracy of the OMRON HEM-7361T automated oscillometric blood pressure (BP) monitor in the differentiation between atrial fibrillation and sinus rhythm.

METHODS

An approximately equal number of patients with persistent atrial fibrillation and individuals with sinus rhythm were recruited from outpatients and inpatients of Ruijin Hospital, Shanghai, China. BP was measured three times consecutively with a 30-s interval with the OMRON HEM-7361T automatic electronic BP monitor for atrial fibrillation detection. A hand-held single lead electrocardiogram device was used for simultaneous electrocardiogram recordings.

RESULTS

The device accurately identified atrial fibrillation in 100 (99.0%) of the 101 patients, with only 1 patient incorrectly classified as non-atrial fibrillation. The device correctly identified 99 (95.2%) of the 104 participants with sinus rhythm as non-atrial fibrillation, with five participants incorrectly classified as atrial fibrillation. The device had a positive predictive value of 95.2%, negative predictive value of 99.0%, and overall accuracy of 97.1%. Among the six misclassified participants, one with atrial fibrillation had a heart rate of 65 beats/min, and four of the five participants with sinus rhythm had cardiac arrhythmias (atrial or ventricular premature beat in one participants, sinus tachycardia in one participant, and both arrhythmias in one participant).

CONCLUSION

The OMRON HEM-7361T BP monitor is accurate in the differentiation between atrial fibrillation and sinus rhythm. Whether the device is sufficiently accurate in the differentiation between atrial fibrillation and other cardiac arrhythmias remains under investigation.

Для цитирования: Zhou Yi, Xia Jia-Hui, Zhang Wei, Li Yan, Wang Ji-Guang. Accuracy of the OMRON HEM-7361T blood pressure monitor in the differentiation between atrial fibrillation and sinus rhythm.. Blood pressure monitoring. 2024. DOI: 10.1097/mbp.0000000000000706.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.