Clinical radiology · 2015
Implementation of a split-bolus single-pass CT protocol at a UK major trauma centre to reduce excess radiation dose in trauma pan-CT.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Leung V, Sastry A, Woo T D, Jones H R
Аннотация
AIM: To quantify the dose reduction and ensure that the use of a split-bolus protocol provided sufficient vascular enhancement.
MATERIALS AND METHODS: Between 1 January 2014 and 31 May 2014, both split bolus and traditional two-phase scans were performed on a single CT scanner (SOMATOM Definition AS+, Siemens Healthcare) using a two-pump injector (Medrad Stellant). Both protocols used Siemens' proprietary tube current and tube voltage modulation techniques (CARE dose and CARE kV). The protocols were compared retrospectively to assess the dose-length product (DLP), aortic radiodensity at the level of the coeliac axis and radiodensity of the portal vein.
RESULTS: There were 151 trauma CT examinations during this period. Seventy-eight used the split-bolus protocol. Seventy-one had traditional two-phase imaging. One patient was excluded as they were under the age of 18 years. The radiodensity measurements for the portal vein were significantly higher (p<0.001) in the split-bolus protocol. The mean aortic enhancement in both protocols exceeded 250 HU, although the traditional two-phase protocol gave greater arterial enhancement (p<0.001) than the split-bolus protocol. The split-bolus protocol had a significantly lower (p<0.001) DLP with 43.5% reduction in the mean DLP compared to the traditional protocol.
CONCLUSION: Split-bolus CT imaging offers significant dose reduction for this relatively young population while retaining both arterial and venous enhancement.
MATERIALS AND METHODS: Between 1 January 2014 and 31 May 2014, both split bolus and traditional two-phase scans were performed on a single CT scanner (SOMATOM Definition AS+, Siemens Healthcare) using a two-pump injector (Medrad Stellant). Both protocols used Siemens' proprietary tube current and tube voltage modulation techniques (CARE dose and CARE kV). The protocols were compared retrospectively to assess the dose-length product (DLP), aortic radiodensity at the level of the coeliac axis and radiodensity of the portal vein.
RESULTS: There were 151 trauma CT examinations during this period. Seventy-eight used the split-bolus protocol. Seventy-one had traditional two-phase imaging. One patient was excluded as they were under the age of 18 years. The radiodensity measurements for the portal vein were significantly higher (p<0.001) in the split-bolus protocol. The mean aortic enhancement in both protocols exceeded 250 HU, although the traditional two-phase protocol gave greater arterial enhancement (p<0.001) than the split-bolus protocol. The split-bolus protocol had a significantly lower (p<0.001) DLP with 43.5% reduction in the mean DLP compared to the traditional protocol.
CONCLUSION: Split-bolus CT imaging offers significant dose reduction for this relatively young population while retaining both arterial and venous enhancement.