Evaluation of a test for its suitability in the diagnosis of variant Creutzfeldt-Jakob disease.
Аннотация
BACKGROUND AND
OBJECTIVES
Evaluation of variant Creutzfeldt-Jakob disease (vCJD) diagnostic/donor screening tests is made complicated by the very limited supply of blood samples from clinically confirmed cases of vCJD. To determine appropriate access for test developers to rare Creutzfeldt-Jakob disease (CJD) blood samples, the oversight committee of the NIBSC CJD Resource Centre has developed a process and protocols detailing minimum requirements for both test sensitivity and specificity. This protocol is broadly similar to that outlined in the common technical specification (European Directive 98/79/EC).
MATERIALS AND
METHODS
Tests are subjected to a stepwise evaluation (step 1). vCJD tissue homogenates spiked into pooled human plasma (step 2). Blood samples from animals known to be incubating (Transmissible spongiform encephalopathy) TSE disease (scrapie/Bovine Spongiform encephalopathy (BSE)-infected sheep, BSE-infected primates) and appropriate controls (step 3). Fresh or frozen plasma from normal UK blood donors and (step 4). Plasma samples from individuals with confirmed clinical stage variant CJD (transfusion transmission) or sporadic CJD (no evidence of blood transmission).
RESULTS
The assay evaluated performed with good sensitivity with vCJD-spiked tissue homogenates, poor sensitivity for ovine TSE-infected blood samples and failed with plasma from BSE-infected non-human primates and with true vCJD clinical samples.
CONCLUSIONS
The test evaluated here is currently unsuitable for use in blood donor screening or diagnosis using blood.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.