Improving Pleural Fluid pH Accuracy: A Quality Improvement Initiative.

Bowmaster Sydney M, Merrill Anna E, Manning Nicole L, Wilson Jeffrey S, Gross Thomas J

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Fluid pH remains a standard test when evaluating pleural effusions. Studies recommend using a blood gas analyzer to measure fluid pH immediately after sample collection. Centralization of laboratory services at our institution led to migration of blood gas analysis to a single core laboratory with uncertain impacts on pH results.

METHODS

Pleural fluid samples were obtained from 30 patients during routine clinical care, excluding those with frank purulence or blood. Fluid pH and pCO2 were measured on a blood gas analyzer. We compared samples collected in plastic screw-top vials from commercially available thoracentesis kits vs paired samples collected in blood gas syringes. We also examined how differences in processing time, fluid pCO2, and heparin anticoagulant correlated with measured pH.

RESULTS

Of the 30 paired samples obtained, 25 (83.3%) were exudates and 5 (16.7%) were transudates. All samples collected in blood gas syringes had lower mean pH and higher mean pCO2 than samples collected in plastic screw-top vials (mean pH difference ± SD, 0.16 ± 0.06; P < 0.001) (mean pCO2 difference ± SD, 14.09 ± 6.92; P < 0.001). Differences in fluid pH were independent of total nucleated cell count (range 65-28 380 cells/mm3). Neither the presence of heparin in the syringe nor time delays up to 2 h between collection and analysis meaningfully impacted the pH result (P = 0.51 and P = 0.75, respectively).

CONCLUSION

To accurately measure pleural fluid pH, specimens should be collected anaerobically in blood gas syringes rather than the screw-top vials provided in thoracentesis kits.

Для цитирования: Bowmaster Sydney M, Merrill Anna E, Manning Nicole L, Wilson Jeffrey S, Gross Thomas J. Improving Pleural Fluid pH Accuracy: A Quality Improvement Initiative.. The journal of applied laboratory medicine. 2025. DOI: 10.1093/jalm/jfaf118.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.