The sensitivity of ECG contamination to surgical implantation site in brain computer interfaces.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Neumann Wolf-Julian, Memarian Sorkhabi Majid, Benjaber Moaad, Feldmann Lucia K, Saryyeva Assel, Krauss Joachim K, Contarino Maria Fiorella, Sieger Tomas, Jech Robert, Tinkhauser Gerd, Pollo Claudio, Palmisano Chiara, Isaias Ioannis U, Cummins Daniel D, Little Simon J, Starr Philip A, Kokkinos Vasileios, Gerd-Helge Schneider, Herrington Todd, Brown Peter, Richardson R Mark, Kühn Andrea A, Denison Timothy

Аннотация

BACKGROUND: Brain sensing devices are approved today for Parkinson's, essential tremor, and epilepsy therapies. Clinical decisions for implants are often influenced by the premise that patients will benefit from using sensing technology. However, artifacts, such as ECG contamination, can render such treatments unreliable. Therefore, clinicians need to understand how surgical decisions may affect artifact probability.

OBJECTIVES: Investigate neural signal contamination with ECG activity in sensing enabled neurostimulation systems, and in particular clinical choices such as implant location that impact signal fidelity.

METHODS: Electric field modeling and empirical signals from 85 patients were used to investigate the relationship between implant location and ECG contamination.

RESULTS: The impact on neural recordings depends on the difference between ECG signal and noise floor of the electrophysiological recording. Empirically, we demonstrate that severe ECG contamination was more than 3.2x higher in left-sided subclavicular implants (48.3%), when compared to right-sided implants (15.3%). Cranial implants did not show ECG contamination.

CONCLUSIONS: Given the relative frequency of corrupted neural signals, we conclude that implant location will impact the ability of brain sensing devices to be used for "closed-loop" algorithms. Clinical adjustments such as implant location can significantly affect signal integrity and need consideration.