Skeletal troponin I cross-reactivity in different cardiac troponin I assay versions.

Hyytiä Heidi, Heikkilä Taina, Hedberg Pirjo, Puolakanaho Tarja, Pettersson Kim

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

To study the skeletal troponin I (skTnI) cross-reactivity of four different commercially available antibodies in four cardiac troponin I (cTnI) research assay versions having the same epitope specificity as evidenced by peptide mapping.

DESIGN AND

METHODS

The four research assays all use two solid phase antibodies and one detection antibody attached to intrinsically fluorescent nanoparticles. Two alternative antibodies were used for one capture antibody and two for the detector antibody. The assays were evaluated in terms of analytical sensitivity and by determining assay cross-reactivity to skTnI. Additionally, regression analysis was performed by measuring a sample panel (n=101) with all of the four assay versions.

RESULTS

A false-positive cTnI concentration of >7000ng/L was measured with one of the assay versions, when serum was spiked with 500,000ng/L skTnI. The corresponding observed cTnI values for the other three assay versions varied from 616ng/L to 727ng/L. Out of the 101 clinical samples assayed, five showed spuriously (3- to 148-fold) elevated cTnI values with the skTnI interference prone assay setup, but not with the other assay versions. According to our investigational skTnI assay, all five samples contained measurable amounts of skTnI (range: 5500-702,000ng/L).

CONCLUSIONS

Two out of four cTnI antibodies tested cross-reacted vastly with skTnI but did not cause any notable interference unless paired together. Therefore, skTnI cross-reactivity should be carefully assessed when cTnI assay antibodies claimed to be cTnI specific are selected.

Для цитирования: Hyytiä Heidi, Heikkilä Taina, Hedberg Pirjo, Puolakanaho Tarja, Pettersson Kim. Skeletal troponin I cross-reactivity in different cardiac troponin I assay versions.. Clinical biochemistry. 2015. DOI: 10.1016/j.clinbiochem.2014.12.028.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.