DOI 10.1002/(sici)1096-9098(199607)62:3<222::aid-jso14>3.0.co;2-8
Catheter fracture: a rare complication of totally implantable subclavian venous access devices.
Аннотация
Catheter fracture represents a rare problem among non-infectious complications following the insertion of totally implantable long-term central venous access systems for the application of chemotherapeutic agents. A literature survey revealed a total incidence of catheter fractures of 0-2.1%. Imminent catheter fracture can be identified radiologically, using different degrees of catheter narrowing between the clavicle and the first rib, called pinch-off sign. Two cases of catheter fracture are described and potential causes are discussed. Recommendations to avoid the pinch-off sign with the subsequent risk of catheter fracture and migration include a more lateral and direct puncture of the subclavian vein. In case of catheter narrowing in the clavicular-first rib angle, patients should be followed carefully by chest X-rays every 4 weeks. Whenever possible, the system should be removed within 6 months following insertion.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.