Differentiating between adductor and abductor spasmodic dysphonia using airflow interruption.
Аннотация
OBJECTIVES/HYPOTHESIS: To measure the laryngeal resistance (R(L)), subglottal pressure (P(s)), and mean flow rate (MFR) of adductor (ADSD) and abductor (ABSD) spasmodic dysphonia patients using the airflow interrupter.
METHODS
The R(L) of six ABSD and seven ADSD patients was measured using the airflow interrupter, a noninvasive device designed to measure MFR and P(s) via mechanical balloon valve interruption. Subjects performed 10 trials at each of two intensity levels, with each trial consisting of a sustained /a/ during which phonation was interrupted for 500 ms. Laryngeal resistance was calculated as subglottal pressure divided by airflow.
RESULTS
Mean R(L) for the ADSD and ABSD subtypes at 65 dB were 24.78 cmH(2)O/L/s and 14.51 cmH(2)O/L/s, respectively (P = .04). Mean R(L) at 70 dB were 40.02 cmH(2)O/L/s and 15.84 cmH(2)O/L/s (P = .014). P(s) for the ADSD and ABSD subtypes at 65 dB were 10.23 cmH(2)O and 8.32 cmH(2)O, respectively (P = .582). At the 70 dB level, P(s) were 12.39 cmH(2)O and 11.78 cmH(2)O (P = .886). MFR for the ADSD and ABSD subtypes at 65 dB were 435 mL/s and 746 mL/s (P = .205). Mean MFR at 70 dB were 518 mL/s and 848 mL/s (P = .198).
CONCLUSIONS
Noninvasive measurements of R(L) may be useful for differentiating between ADSD and ABSD. This simple objective test, which produces a quantitative output, could be used to evaluate laryngeal function in patients with spasmodic dysphonia.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.