DOI 10.1016/j.clinbiochem.2014.12.015
Analytical sensitivity of four commonly used hCG point of care devices.
Аннотация
BACKGROUND
Point of care (POC) hCG assays are often used to rule-out pregnancy and therefore diagnostic sensitivity, especially at low concentrations of hCG, is important. There are very few studies in the literature that seek to verify the claimed analytical sensitivity of hCG POC devices.
METHODS
The analytical sensitivity of four commonly used hCG POC devices (Alere hCG Combo Cassette, ICON 20 hCG, OSOM hCG Combo Test, and Sure-Vue Serum/Urine hCG-STAT) was challenged using urine samples (n=50) selected based on quantitative hCG concentrations. The majority of these specimens (n=40) had an hCG concentration between 20 and 200 U/L. Each specimen/device combination was reviewed by three individuals. Statistical calculations were performed using Stata 12.
RESULTS
The analytical sensitivity of the OSOM was significantly lower inferior than that of the other POC devices. There was no significant difference in the sensitivity of the Alere, ICON 20 and Sure-Vue devices. There was no significant difference in the individual interpretation of the hCG POC results.
CONCLUSIONS
All hCG POC devices evaluated in this study were susceptible to false negative results at low concentrations of urine hCG. Laboratorians and clinicians should be aware that there are limitations when using urine hCG POC devices to rule out early pregnancy.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.