The mechanisms of positional dysfunction of subclavian venous catheters.

Krutchen A E, Bjarnason H, Stackhouse D J, Nazarian G K, Magney J E, Hunter D W

СтатьяПеревод в очереди

Аннотация

PURPOSE: To determine the origin of subclavian vein catheter and lead dysfunction.

MATERIALS AND

METHODS

Cineradiography was performed on 10 patients with subclavian venous catheter dysfunction and three patients with pacemaker or defibrillator lead dysfunction. The leads and catheters were removed and replaced with use of a fluoroscopically guided technique; the needle entered the vein lateral to the first rib. Repeat cine examinations were performed following placement of new catheters.

RESULTS

The cause of the dysfunction of all 10 catheters was shown to be pinch by the subclavicular musculotendinous tissues as the catheter passed below the clavicle toward its entry into the vein. All three leads were entrapped in the subclavicular tissues and stretched during abduction. The abnormal motion and clinical problems were eliminated after replacement.

CONCLUSION

Subclavian catheter and lead malfunction is not due to compression between the first rib and the clavicle. It is due to entrapment in the subclavius muscle-costoclavicular ligament complex, which binds or compresses the device during movements. These problems can be avoided by employing fluoroscopically guided puncture techniques that enter the vein lateral to the first rib.

Для цитирования: Krutchen A E, Bjarnason H, Stackhouse D J, Nazarian G K, Magney J E, Hunter D W. The mechanisms of positional dysfunction of subclavian venous catheters.. Radiology. 1996. DOI: 10.1148/radiology.200.1.8657904.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.