Effect of thin overlapping reconstruction on the attenuation of small (< or = 3 cm) renal cysts in the nephrographic phase of MDCT: a phantom study.

Kim Sun Ho, Kim Seung Hyup

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To evaluate the effect of thin overlapping reconstruction on the attenuation of small (< or =3 cm) renal cysts in the nephrographic phase of multidetector CT (MDCT).

MATERIALS AND

METHODS

We scanned a phantom kidney containing spheres of various sizes (10, 20, and 30 mm) using both 4- and 16-channel MDCT scanners, and reconstructed images with various slice thickness (T, mm) and intervals (I, mm). The attenuation increase (AI) was measured for each sphere in 240-HU diluted solution of contrast material and compared with the attenuation in 35-HU solution.

RESULTS

On the 4-channel MDCT, thin overlapping reconstruction (T/I = 3/1, compared with 5/5) lowered the AI as much as 17 HU in the 10 mm-sphere and 6 HU in the 20 mm-sphere (p < 0.05). Thin slicing alone was also effective; however overlapping alone was not. On the 16-channel MDCT, AI in the 10 mm-sphere was significantly lower than on the 4-channel MDCT with T/I = 5/5 (p < 0.05), however thinner slicing or overlapping did not affect the attenuation significantly in all of the spheres.

CONCLUSION

The effect of thin overlapping reconstruction on minimizing falsely elevated attenuation in the nephrographic phase was significant only in cysts < or = 20 mm on the 4-channel MDCT.

Для цитирования: Kim Sun Ho, Kim Seung Hyup. Effect of thin overlapping reconstruction on the attenuation of small (< or = 3 cm) renal cysts in the nephrographic phase of MDCT: a phantom study.. Korean journal of radiology. 2006. DOI: 10.3348/kjr.2006.7.4.275.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.