DOI 10.1016/j.ajic.2018.04.215
Evaluation of adenosine triphosphate test for cleaning assessment of gastroscopes and the effect on workload in a busy endoscopy center.
Аннотация
OBJECTIVE
Using adenosine triphosphate (ATP) tests to assess manual cleaning of gastroscopes and to determine the associated workload in a busy endoscopy unit.
METHODS
Patient-used gastroscopes were sampled before and after cleaning to assess ATP levels, bioburden, and protein. Samples were collected by flushing 20 mL of sterile water through the biopsy port to the distal end. Time spent for reprocessing and performing the ATP test was recorded.
RESULTS
Twenty-four samples were collected from 10 gastroscopes. After manual cleaning, 14/24 (58.3%) samples had no microbial growth (mean, 21 colony-forming units/cm2), and in 22/24 (91.7%) samples the protein was undetectable (mean, 0.04 µg/cm2). ATP test was above the cutoff (200 relative light units [RLU]) in 17/24 (70.8%) samples (mean, 498 RLU). After the second cleaning, 11/17 (64.7%) gastroscopes still failed the ATP test (mean, 321.2 RLU). The mean time spent to perform manual cleaning and ATP tests was 16 and 8 minutes, respectively. Hence, each test increased the length of time for cleaning plus testing cleanliness by 50%.
CONCLUSION
Further studies regarding the optimal cutoff for ATP tests are needed. ATP tests for cleaning monitoring are easy to perform and provide immediate feedback to the team. However, the increased workload needs to be considered.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.