MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/0007-1226(94)90030-2

Медицинская библиотека · British journal of plastic surgery · 1994

The role of ultrasound imaging in the diagnosis of breast implant rupture: a prospective study.

Liston J C, Malata C M, Varma S, Scott M, Sharpe D T

СтатьяПеревод в очереди

Аннотация

Breast ultrasound to assess possible implant rupture was performed on 24 consecutive patients (43 breasts) by the same radiologist immediately before revisional breast implant surgery. Comparison of the clinical, ultrasound and operative findings showed ultrasound to be three times as sensitive as clinical judgement in predicting implant rupture. The sensitivity of ultrasound in predicting impaired implant integrity was 70% (versus 23% for clinical examination) with a specificity of 96%. The ultrasound features indicating leakage are outlined later. The positive predictive value of an abnormal scan was 90% and that of a normal scan 87%. It is concluded that breast ultrasound is a simple, quick, non-invasive method which contributes significantly to the assessment of patients with suspected breast implant rupture.

Для цитирования: Liston J C, Malata C M, Varma S, Scott M, Sharpe D T. The role of ultrasound imaging in the diagnosis of breast implant rupture: a prospective study.. British journal of plastic surgery. 1994. DOI: 10.1016/0007-1226(94)90030-2.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.