DDD and single-lead VDD pacing: evaluation of atrial signal dynamic changes.

Оценочное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Boriani G, Biffi M, Bandini A, Moracchini P V, Capucci A, Sigliano R, Mininno A, Branzi A

Аннотация

BACKGROUND: Single-lead VDD pacing systems are an alternative to conventional DDD pacemakers in patients with atrioventricular (AV) block and normal sinus function.

HYPOTHESIS: The aim of this study was to assess changes of P-wave amplitude occurring in dynamic conditions in two groups of patients with a single-lead VDD and with a DDD pacing system, respectively.

METHODS: Twenty-eight patients with second- or third-degree AV block and normal sinus function were enrolled prospectively into the study. Seventeen patients were implanted with a single-lead VDD pacing system and 11 with a DDD pacemaker. Patients were evaluated at 3 months (all patients) and at 6 months (26 patients) at supine and in dynamic conditions (postural changes, hyperventilation, and during exercise).

RESULTS: Mean P-wave values at supine were 1.92 +/- 1.10 mV at 3 months and 1.76 +/- 1.01 mV at 6 months for VDD systems, and 4.63 +/- 2.18 mV at 3 months and 4.58 +/- 2.80 mV at 6 months for DDD pacemakers. In dynamic conditions, P-wave amplitude changes compared with supine condition ranged between -74 and +226% in VDD, and between -53 and +138% in DDD; however P-wave amplitudes showed no significant changes compared with baseline. Moreover, changes in atrial signal amplitudes did not occur randomly, and in both systems P-wave amplitudes remained significantly correlated with supine values.

CONCLUSIONS: A wide range of P-wave amplitude variations occurs in different postural conditions or during exercise, both with single-lead VDD and DDD pacing systems. However, with appropriate programming of atrial sensitivity based on supine values, constant atrial tracking can be maintained.