DOI 10.1007/s11255-023-03756-1
Using a stress reservoir to improve urine leakage after artificial urinary sphincter implantation.
Аннотация
INTRODUCTION
Although the longest efficacy record, some patients report about urinary leakage during higher intra-abdominal pressure after Artificial Urinary Sphincter (AUS) implantation. To improve the continent results, we placed in addition to the occluding cuff and the pressure regulating balloon, in a second procedure a stress-relief reservoir in lower abdomen, for additional passive pressure transmission to the occluding cuff.
METHODS
In this retrospective monocentric data analysis between 2011 and 2018, 80 patients with persistence incontinence after AUS implantation were included. Stress-relief reservoir was indicated in 12 patients with involuntary leakage of urine, that occurred when intra-abdominal pressure raised.
RESULTS
In all 12 cases, the stress reservoir was easily implanted and there were no intraoperative complications. In a mean follow-up time of 53 months, the pad per day usage (p/d) improved from 3 (± 1.2) to 1.7 (± 1.5) (p = 0.001). Two patients with multiple previous abdominal surgeries used an equal number of pads after SRR; however, an improvement during physical exertion was reported. Continence situation was evaluated with a questionnaire and was rated as "excellent", "good", or "satisfactory" by 11 (92%) patients.
CONCLUSION
The persistence of urinary incontinence after AUS is a challenging topic. Implantation of a stress reservoir in carefully selected patients with urinary leakage during higher intra-abdominal pressure is minimally invasive and offers new options to improve the proven long-term record of AUS. Certainly, more investigations are needed to determine the clinical relevance of this approach.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.