DOI 10.1080/00365513.2023.2255970
The performance of saliva test strips for determining ethanol levels, as compared to gas chromatography and breathalyser methods.
Аннотация
We set out to determine the performance of the Testi Technologies enzymatic assay saliva ethanol test strips of three different detection levels: 0 g/L, 0.2 g/L and 0.5 g/L, using as the reference method a gas chromatography analyser (GC). Alcohol levels were measured in 104 volunteers at up to three points in time, using up to three test strips per measurement, while gathering blood samples and breathalyser readings in parallel. The plasma alcohol concentrations (PAC) were determined from the plasma samples using GC. The qualitative results of the test strips were compared to the quantitative results from the reference method, as well as the breathalyser readings, and the amount of true and false positive and true and false negative results were classified using predetermined cut-off levels. The best performing test strips were the 0 g/L and the 0.2 g/L strips. The 0 g/L strips had a sensitivity and specificity of 1.00, as false negatives and false positives were not detected. The 0.2 g/L strips had a sensitivity and specificity [95% confidence interval (CI)] of 0.98 (0.96 - 1.00) and 0.83 (0.62 - 1.00) respectively, an accuracy of 0.97 (0.95 - 0.99), and a diagnostic odds ratio of 205.00 (35.33 - 1189.66). The test strips perform their intended purpose of screening for alcohol consumption well, with their great sensitivity as a defining property compared to other testing methods. For them to be able to be implemented in a clinical setting however, further refinement of the tests' characteristics would be required.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.