Validation of the eyeTelemed IOPvet indentation tonometer for use in dogs.

Kapeller Lydia E, Cabble Ava G, Buckman Phillip N, Harman Christine D, Jacobson Amanda L, Lawrence Frank R, Komáromy András M

СтатьяВалидационное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

To assess the accuracy of canine intraocular pressure (IOP) estimates from the eyeTelemed IOPvet indentation tonometer.

ANIMALS STUDIED: Part 1 included 54 eyes from 28 Beagle dogs-23 ADAMTS10-mutants with open-angle glaucoma and 5 normals. Part 2 involved five normal canine ex vivo globes.

PROCEDURE: Part 1 (in vivo) compared IOPvet estimates in normal and glaucomatous dogs to Reichert Tono-Vera® Vet rebound tonometry. The three IOPvet estimates were green (normal; <20 mmHg, according to the manufacturer), yellow (elevated; 20-30 mmHg), and red (high; >30 mmHg). In Part 2 (ex vivo), the pressure inside freshly enucleated normal canine eyes was progressively increased from 5 to 80 mmHg and compared to IOPvet estimates. Descriptive statistics compared IOPvet estimates to rebound tonometry and direct manometry, with the threshold from normal to glaucoma set at 30 mmHg.

RESULTS

In Part 1 (in vivo), normal pressures (≤30 mmHg) were mainly identified correctly as green or yellow-110 of 111 estimates, corresponding to a specificity of 99%. Only 16 of 125 affected estimates were correctly displayed in the >30-mmHg range; the remaining 109 showed ≤30 mmHg, corresponding to a sensitivity of 13%. In Part 2 (ex vivo), all normal pressures were correctly estimated with green, but 64 of 88 manometric IOPs >30 mmHg were falsely estimated as 20-30 mmHg.

CONCLUSIONS

The IOPvet is inaccurate in estimating canine IOP with a low sensitivity at identifying dogs with IOP > 30 mmHg. Canine-specific instrument revision is required to correctly identify elevated (yellow = 20-30 mmHg) and high (red >30 mmHg) IOPs.

Для цитирования: Kapeller Lydia E, Cabble Ava G, Buckman Phillip N, Harman Christine D, Jacobson Amanda L, Lawrence Frank R, Komáromy András M. Validation of the eyeTelemed IOPvet indentation tonometer for use in dogs.. Veterinary ophthalmology. 2024. DOI: 10.1111/vop.13215.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.