Premature ventricular complexes: Assessing burden density in a large national cohort to better define optimal ECG monitoring duration.

Krumerman Andrew, Di Biase Luigi, Gerstenfeld Edward, Dickfeld Timm, Verma Nishant, Liberman Leonardo, Amara Richard, Kacorri Ardit, Crosson Lori, Wilk Alan, Ferrick Kevin J

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Premature ventricular contraction (PVC) burden is a risk factor for heart failure and cardiovascular death in patients with structural heart disease. Long-term electrocardiographic monitoring can have a significant impact on PVC burden evaluation by further defining PVC distribution patterns.

OBJECTIVE

This study aimed to ascertain the optimal duration of electrocardiographic monitoring to characterize PVC burden and to understand clinical characteristics associated with frequent PVCs and nonsustained ventricular tachycardia in a large US cohort.

METHODS

Commercial data (iRhythm's Zio patch) from June 2011 to April 2022 were analyzed. Inclusion criteria were age >18 years, PVC burden ≥5%, and wear period ≥13 days. PVC burden cutoffs were determined on the basis of AHA/ACC/HRS guidelines for very frequent PVCs (10,000-20,000 during 24 hours). Patients were assigned to categories by PVC densities: low, <10%; moderate, 10% to <20%; and high, ≥20%. Mean measured error was assessed at baseline and daily until the wear period's end for overall PVC burden and different PVC densities.

RESULTS

Analysis of 106,705 patch monitors revealed a study population with mean age of 70.6 ± 14.6 years (33.6% female). PVC burden was higher in male patients and those >65 years of age. PVC burden mean error decreased from 2.9% at 24 hours to 1.3% at 7 days and 0.7% at 10 days. Number of ventricular tachycardia episodes per patient increased with increasing PVC burden (P < .0001).

CONCLUSION

Extending ambulatory monitoring beyond 24 hours to 7 days or more improves accuracy of assessing PVC burden. Ventricular tachycardia frequency and duration vary by initial PVC density, highlighting the need for prolonged cardiac monitoring.

Для цитирования: Krumerman Andrew, Di Biase Luigi, Gerstenfeld Edward, Dickfeld Timm, Verma Nishant, Liberman Leonardo, Amara Richard, Kacorri Ardit, Crosson Lori, Wilk Alan, Ferrick Kevin J. Premature ventricular complexes: Assessing burden density in a large national cohort to better define optimal ECG monitoring duration.. Heart rhythm. 2024. DOI: 10.1016/j.hrthm.2024.04.066.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.