Performance Verification of the Iris iQ200 Sprint Automated Urine Microscopy Analyzer in a Hospital Routine Laboratory.

Cui Ming, Ju Shaoqing, Shi Yingjuan, Jing Rongrong

Оценочное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Systematic performance verification is required before a laboratory can introduce a new measurement procedure for reporting results of patient testing. The aim of this study was to determine whether a new Iris iQ200 Sprint automated urine microscopy analyzer (iQ200 Sprint) could be incorporated into our routine laboratory.

METHODS

A total of 421 fresh urine samples were selected from the Affiliated Hospital of Nantong University, including those from healthy individuals and those with a variety of abnormalities to ensure a wide range of results. Precision, recovery, carry-over, linearity and reference interval were verified according to well-established protocols.

RESULTS

The repeatability studies found coefficients of variability (CVs) in the range of 10.53% - 20.28% for red blood cells, white blood cells, and squamous epithelial cells, while the CV for the iQ Positive Control sample was 3.23%. The relative bias was 0.5% for the iQ Positive Control sample and no carry-over was detected. Linearity was observed at concentrations of 10 - 2069.5 particles/μL (y = 0.989x + 9.1, R2 = 0.999). The manufacturer's claimed reference interval meets the requirements for medical usefulness.

CONCLUSIONS

Performance verification is needed before a clinical laboratory can introduce a new measurement procedure. The iQ200 Sprint is sufficiently precise and reliable to be applied in our clinical laboratory.

Для цитирования: Cui Ming, Ju Shaoqing, Shi Yingjuan, Jing Rongrong. Performance Verification of the Iris iQ200 Sprint Automated Urine Microscopy Analyzer in a Hospital Routine Laboratory.. Clinical laboratory. 2017. DOI: 10.7754/clin.lab.2017.170318.