DOI 10.1053/j.jvca.2023.08.004
Continuous Noninvasive Blood Pressure Measurement With "ClearSight" Compared to Standard Intermittent Blood Pressure Measurement in Patients With Peripheral Arterial Disease. Are Potential Differences Influenced by Phenylephrine or Dobutamine?
Аннотация
OBJECTIVES
To investigate the agreement between continuous noninvasive blood pressure measurement with the ClearSight system (cNIBP-CS) and standard intermittent noninvasive blood pressure measurement (iNIBP) in patients with peripheral arterial disease (PAD). Additionally, the influence of vasoactive medication on potential measurement differences was assessed.
DESIGN
A secondary analysis of a randomized controlled trial.
SETTING
At a university hospital.
PARTICIPANTS
Thirty-four patients with PAD undergoing percutaneous transluminal angioplasty of the lower limbs.
INTERVENTIONS: None.
MEASUREMENTS AND MAIN
RESULTS
Continuous noninvasive blood pressures were measured with the "ClearSight" system and compared to standard iNIBPs. Bland-Altman analysis revealed a mean bias of 13 mmHg (±15) between cNIBP-CS and iNIBP, with 95% limits of agreement (LOA) ranging from -17 to 42 mmHg. When comparing both medication groups, a similar mean bias was found for phenylephrine and dobutamine (12 mmHg [±13] and 13 mmHg [±13], respectively).
CONCLUSION
In this study, in patients with PAD, cNIBP-CS showed an underestimation of blood pressure compared to iNIBP in phenylephrine- and dobutamine-treated patients. Compared to previous studies, a larger bias and wider 95% LOA were found.