MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1161/circulationaha.109.929810

Медицинская библиотека · Circulation · 2010

Visceral arterial compromise during intra-aortic balloon counterpulsation therapy.

Rastan Ardawan Julian, Tillmann Eugen, Subramanian Sreekumar, Lehmkuhl Lukas, Funkat Anne Kathrin, Leontyev Sergej, Doenst Torsten, Walther Thomas, Gutberlet Matthias, Mohr Friedrich Wilhelm

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Intraaortic balloon pump (IABP) therapy is a widely used method of circulatory support. Based on frequent findings of balloon malposition with visceral arterial compromise on computerized tomographic (CT) imaging studies, we conducted a systematic review of cardiac surgical IABP patients with available CT scans to determine reasons, incidence, and clinical relevance of malposition.

METHODS AND

RESULTS

From January 2007 to March 2009, a total of 621 of 7756 cardiac surgical patients (8.0%) received perioperative IABP support, of whom 63 (10.1%) received a thoracoabdominal CT scan during IABP support. Proximal and distal balloon positions were analyzed. The anatomic distance from the left subclavian artery to celiac trunk and aortic transverse diameter were measured in all patients and compared with implanted balloon dimensions. Mean age was 67.1±11.9 years; 33.3% were female, and height was 169±9 cm. Based on radiography, proximal balloon position was correct in 96.8% but only appropriate in 38.1% based on CT. In 61 of 63 patients, compromise of at least 1 visceral artery was found: celiac trunk, 96.8%; superior mesenteric artery, 87.3%; and renal arteries, 66.7%. Left subclavian artery to celiac trunk distance was 241±23 mm, and balloon length was 248±17 mm and corresponded to an anatomic to balloon length mismatch in 68.2%. Spinal deformations were found in 42.9%. Laparotomy for mesenteric ischemia was required in 23.8%. Hospital mortality rate was 60.3%.

CONCLUSIONS

IABP malposition was commonly identified by CT. Reasons included incorrect proximal balloon position as well as an anatomic-to-balloon length mismatch. Thus, shorter than recommended balloon sizes and better positioning strategies had to be considered.

Для цитирования: Rastan Ardawan Julian, Tillmann Eugen, Subramanian Sreekumar, Lehmkuhl Lukas, Funkat Anne Kathrin, Leontyev Sergej, Doenst Torsten, Walther Thomas, Gutberlet Matthias, Mohr Friedrich Wilhelm. Visceral arterial compromise during intra-aortic balloon counterpulsation therapy.. Circulation. 2010. DOI: 10.1161/circulationaha.109.929810.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление