DOI 10.1111/ijlh.12723

Медицинская библиотека · International journal of laboratory hematology · 2017

Two-site evaluation of the diagnostic performance of the Sysmex XN Body Fluid (BF) module for cell count and differential in Cerebrospinal Fluid.

Buoro S, Peruzzi B, Fanelli A, Seghezzi M, Manenti B, Lorubbio M, Biagioli T, Nannini S, Ottomano C, Lippi G

Оценочное исследованиеСтатьяПеревод в очереди

Аннотация

INTRODUCTION

Cellular analysis in cerebrospinal fluid (CSF) provides important diagnostic information in many pathological settings. The aim of this two-site study was to evaluate the Sysmex XN Body Fluid mode (XN-BF) for cell analysis of CSF compared to light microscopy (LM).

METHODS

Two hundred and seven consecutive CSF samples were analyzed in parallel with XN-BF and LM. The study also included the estimation of the limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ), carry-over and linearity of XN-BF module.

RESULTS

LoQ of white blood cells (WBC) was 3×106  cells/L; linearity was good and carry-over negligible. XN-BF parameters were compared to LM for the following cell classes: total cells, WBC, polymorphonuclear (PMN), and mononuclear (MN) cells. The bias ranged from 1.3 to 15.2×106  cells/L. The receiver operating characteristics curve analysis for WBC showed an area under the curve of 0.98, and the global diagnostic agreement was 95% at a cutoff of 5×106  cells/L.

CONCLUSIONS

XN-BF provides rapid and accurate counts in clinically relevant ranges of CSF values, thus providing a valuable alternative to conventional LM analysis. However, microscopic review remains advisable in samples with abnormal cell counts or high fluorescent (HF-BF) cell parameter exceeding 5×106  cells/L.

Для цитирования: Buoro S, Peruzzi B, Fanelli A, Seghezzi M, Manenti B, Lorubbio M, Biagioli T, Nannini S, Ottomano C, Lippi G. Two-site evaluation of the diagnostic performance of the Sysmex XN Body Fluid (BF) module for cell count and differential in Cerebrospinal Fluid.. International journal of laboratory hematology. 2017. DOI: 10.1111/ijlh.12723.

Предметные рубрики MeSH