Criterion (Concurrent) Validity and Clinical Utility of the Tongueometer Device.

Krekeler Brittany N, Hopkins Anna, Tabangin Meredith E, Altaye Mekibib, Roberts Rachel, Saadi Raneh, Martin-Harris Bonnie, Rogus-Pulia Nicole

СтатьяВалидационное исследованиеПеревод в очереди

Аннотация

PURPOSE: Tongue manometry (i.e., tongue pressure measurement) is a commonly used assessment for patients with suspected oral-motor involvement in swallowing disorders. Availability of lingual manometry has changed in recent years, with the introduction of the Tongueometer device being a more affordable tongue manometry system. The purpose of this study was to test concurrent (criterion) validity of the Tongueometer compared to the current standard reference device, the Iowa Oral Performance Instrument (IOPI).

METHOD: Adults without dysphagia were recruited for participation in this study. Standard lingual measurements (swallowing-related pressures, maximum isometric pressure [MIP], and maximum isometric endurance) were recorded, with the bulb anteriorly placed, with both devices, in a randomized order. The Bland-Altman method was used to determine concurrent (criterion) validity of these measurements compared to the clinical standard IOPI device. A recently available suggested corrective value by Curtis et al. (2023) was also applied, with comparisons made between devices both with and without the Curtis correction.

RESULTS

The final sample included 70 adult participants aged 20-89 years (Mage = 52.3 years). Measures with the Tongueometer device were significantly lower when compared with the same measures taken using the IOPI (p < .01) for all measures including MIP, endurance, and swallow pressures. The correction suggested by Curtis and colleagues did not ameliorate these differences.

CONCLUSIONS

The Tongueometer lingual measurements were consistently lower compared to the IOPI. Clinical use of values taken with the Tongueometer device should be compared to normative data published for each specific device. Available features of each device (e.g., display, bulb texture, technology/application) should be considered when selecting which device to use with an individual patient.

Для цитирования: Krekeler Brittany N, Hopkins Anna, Tabangin Meredith E, Altaye Mekibib, Roberts Rachel, Saadi Raneh, Martin-Harris Bonnie, Rogus-Pulia Nicole. Criterion (Concurrent) Validity and Clinical Utility of the Tongueometer Device.. American journal of speech-language pathology. 2024. DOI: 10.1044/2024_ajslp-23-00478.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.