Transcatheter closure of moderate-to-large patent ductus arteriosus in infants using Amplatzer duct occluder.
Аннотация
BACKGROUND
There are difficulties in transcatheter closure of patent ductus arteriosus (PDA) in infants.
METHODS AND
RESULTS
The 46 infants (mean age 6.2+/-2.7 months; mean body weight 6.3+/-1.6 kg) who underwent PDA closure using the Amplatzer duct occluder (ADO). The indication for using an ADO was a ductus diameter > or =2.5 or 3 mm. Device diameter selected was 1-3 mm larger than ductal diameter. The mean systolic pulmonary artery pressure was 40.9+/-18.2 mmHg. The mean Qp/Qs ratio was 3.1+/-1.2. The mean ductus diameter was 3.3+/-0.8 mm. ADO was successfully deployed in 45 patients. Failure occurred in 1 case. The mean diameter of device used was 5.4+/-1.1 mm. No severe complications occurred. At the 1-month echocardiographic follow-up, a small residual shunt was present in 4 of 45 patients and had disappeared in all 4 patients at the 3-month follow-up. One patient developed a moderate degree of left ventricular outflow tract obstruction 2.3 years after the procedure.
CONCLUSIONS
Transcatheter closure of PDA in infants using the ADO is a safe and effective method.