DOI 10.1253/circj.cj-09-0473

Медицинская библиотека · Circulation journal : official journal of the Japanese Circulation Society · 2009

Transcatheter closure of moderate-to-large patent ductus arteriosus in infants using Amplatzer duct occluder.

Wang Jou-Kou, Wu Mei-Hwan, Lin Ming-Tai, Chiu Shuenn-Nan, Chen Chun-An, Chiu Hsin-Hui

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

There are difficulties in transcatheter closure of patent ductus arteriosus (PDA) in infants.

METHODS AND

RESULTS

The 46 infants (mean age 6.2+/-2.7 months; mean body weight 6.3+/-1.6 kg) who underwent PDA closure using the Amplatzer duct occluder (ADO). The indication for using an ADO was a ductus diameter > or =2.5 or 3 mm. Device diameter selected was 1-3 mm larger than ductal diameter. The mean systolic pulmonary artery pressure was 40.9+/-18.2 mmHg. The mean Qp/Qs ratio was 3.1+/-1.2. The mean ductus diameter was 3.3+/-0.8 mm. ADO was successfully deployed in 45 patients. Failure occurred in 1 case. The mean diameter of device used was 5.4+/-1.1 mm. No severe complications occurred. At the 1-month echocardiographic follow-up, a small residual shunt was present in 4 of 45 patients and had disappeared in all 4 patients at the 3-month follow-up. One patient developed a moderate degree of left ventricular outflow tract obstruction 2.3 years after the procedure.

CONCLUSIONS

Transcatheter closure of PDA in infants using the ADO is a safe and effective method.

Для цитирования: Wang Jou-Kou, Wu Mei-Hwan, Lin Ming-Tai, Chiu Shuenn-Nan, Chen Chun-An, Chiu Hsin-Hui. Transcatheter closure of moderate-to-large patent ductus arteriosus in infants using Amplatzer duct occluder.. Circulation journal : official journal of the Japanese Circulation Society. 2009. DOI: 10.1253/circj.cj-09-0473.