Clinica chimica acta; international journal of clinical chemistry · 2017
Preliminary evaluation of UF-5000 Body Fluid Mode for automated cerebrospinal fluid cell counting.
Оценочное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Seghezzi Michela, Manenti Barbara, Previtali Giulia, Alessio Maria Grazia, Dominoni Paola, Buoro Sabrina
Аннотация
BACKGROUND: Cellular analysis of cerebrospinal fluid (CSF) provides important diagnostic information in various medical conditions. The aim of this study was to evaluate the application of Sysmex UF-5000 body fluid mode in cytometric analysis of CSF compared to Light Microscopic (LM).
METHODS: Eighty-one consecutive CSF samples were analyzed by UF-5000 body fluid mode and by LM. The study also included the evaluation of: limit of Blank (LoB), limit of Detection (LoD), limit of Quantitation (LoQ), carryover and linearity.
RESULTS: For total nucleated cells (TNC-UF) and white blood cells (WBC-UF) LoB, LoD and LoQ were 1×106cells/L, 1.8×106cells/L and 1.9×106cells/L respectively. For red blood cells (RBC) LoB was 2×106cells/L, LoD was 3.5×106cells/L and LoQ was 14×106cells/L respectively. Linearity was excellent, carryover was negligible. The agreement between UF-5000 body fluid mode parameters and manual cell counts was good in all CSF samples with bias ranged between -0.5 and 25.1×106cells/L. The ROC curve analysis showed an area under curve of 0.99 for both TNC-UF and WBC-UF parameters.
CONCLUSIONS: The UF-5000 body fluid mode offers rapid and accurate counts in clinically relevant concentration ranges, replacing the LM for most samples. However, in samples with abnormal cell counts or with abnormal scattergram the need for microscopic review remains.
METHODS: Eighty-one consecutive CSF samples were analyzed by UF-5000 body fluid mode and by LM. The study also included the evaluation of: limit of Blank (LoB), limit of Detection (LoD), limit of Quantitation (LoQ), carryover and linearity.
RESULTS: For total nucleated cells (TNC-UF) and white blood cells (WBC-UF) LoB, LoD and LoQ were 1×106cells/L, 1.8×106cells/L and 1.9×106cells/L respectively. For red blood cells (RBC) LoB was 2×106cells/L, LoD was 3.5×106cells/L and LoQ was 14×106cells/L respectively. Linearity was excellent, carryover was negligible. The agreement between UF-5000 body fluid mode parameters and manual cell counts was good in all CSF samples with bias ranged between -0.5 and 25.1×106cells/L. The ROC curve analysis showed an area under curve of 0.99 for both TNC-UF and WBC-UF parameters.
CONCLUSIONS: The UF-5000 body fluid mode offers rapid and accurate counts in clinically relevant concentration ranges, replacing the LM for most samples. However, in samples with abnormal cell counts or with abnormal scattergram the need for microscopic review remains.