Evaluation of the validity and reliability of A-scan ultrasound biometry with a single use disposable cover.

Cass K, Thompson C M, Tromans C, Wood I C J

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The UK Medical Devices Agency has suggested that ophthalmic practitioners should, where practicable and not compromising clinical outcome, restrict corneal contact devices to single patient use to minimise a remote theoretical risk of transmission of new variant Creutzfeldt-Jakob disease (vCJD). This study reports on a modified technique of ultrasound A-scan biometry that complies with the MDA recommendations.

METHODS

The right eyes of 37 consecutive hospital patients had a series of biometry readings taken with a Humphrey 820 A-scan instrument with a plane wave transducer use d conventionally and with the addition of a disposable latex cover.

RESULTS

Intrasessional repeatability of axial length measurements was similar for conventional readings--mean difference 0.027 mm, 95% confidence intervals (CI) +/- 0.44 mm and those taken with a disposable cover (0.028 mm, CI +/- 0.38). Intersessional repeatability was equivalent with (0.002 mm, CI +.- 0.51) and without a cover (0.03 mm, CI +/- 0.51). Readings with a cover were not significantly different from those without (paired t test; p >0.05), but tended to be greater (mean difference 0.085 mm, CI +/- 0.60).

CONCLUSIONS

These findings suggest that corneal contact biometry with a disposable cover is a viable and theoretically safer alternative to the conventional technique.

Для цитирования: Cass K, Thompson C M, Tromans C, Wood I C J. Evaluation of the validity and reliability of A-scan ultrasound biometry with a single use disposable cover.. The British journal of ophthalmology. 2002. DOI: 10.1136/bjo.86.3.344.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.