DOI 10.1007/s12350-014-9963-8

Медицинская библиотека · Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology · 2014

Data-driven respiratory motion tracking and compensation in CZT cameras: a comprehensive analysis of phantom and human images.

Ko Chi-Lun, Wu Yen-Wen, Cheng Mei-Fang, Yen Ruoh-Fang, Wu Wen-Chau, Tzen Kai-Yuan

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

This study described a method for tracking and compensating respiratory motion in cadmium-zinc-telluride (CZT) cameras. We evaluated motion effects on myocardial perfusion imaging and assessed the usefulness of motion compensation in phantom and clinical studies.

METHODS

SPECT studies were obtained from an oscillating heart phantom and 552 patients using CZT cameras with list-mode acquisition. Images were reformatted in 500-ms frames, and the activity centroid was calculated as respiratory signal. The myocardial perfusion, left ventricular (LV) wall thickness, and LV volume were assessed before and after the motion compensation technique.

RESULTS

In phantom studies, we documented only minimal bias between simulated and measured shifts. Significantly reduced tracer activity, increased wall thickness and decreased volume in scans with 15 mm or more axial shifts were noted. In clinical studies, there was a higher prevalence of significant motion after treadmill exercise. The motion compensation technique could successfully compensate those motion artifacts.

CONCLUSION

The described method allows for tracking and compensating respiratory motion in CZT cameras. Significant respiratory motion is still not uncommon using CZT cameras, especially in patients who underwent treadmill tests. Motion blurring can be compensated using image processing techniques and image quality could be significantly improved.

Для цитирования: Ko Chi-Lun, Wu Yen-Wen, Cheng Mei-Fang, Yen Ruoh-Fang, Wu Wen-Chau, Tzen Kai-Yuan. Data-driven respiratory motion tracking and compensation in CZT cameras: a comprehensive analysis of phantom and human images.. Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology. 2014. DOI: 10.1007/s12350-014-9963-8.