Evaluation of 24-second cyanide-containing and cyanide-free methods for whole blood hemoglobin on the Technicon H*1TM analyzer with normal and abnormal blood samples.

Malin M J, Sclafani L D, Wyatt J L

СтатьяПеревод в очереди

Аннотация

Automated methods, with and without cyanide (+CN and -CN), for whole blood hemoglobin (Hb) determination were evaluated on the Technicon H*1TM System. Both automated Hb methods were linear over the range 0-250 g/L (0-25 g/dL) and correlated well with the International Committee for Standardization of Hematology (ICSH) reference method and with the Coulter S+II. Both methods quantitatively converted whole blood containing up to 100% carboxyhemoglobin in less than 24 seconds to their respective end products. With respect to abnormal samples (sickle cell anemia, multiple myeloma, and hyperlipemia), both H*1 methods gave Hb results that were equivalent to the (postfiltration) ICSH method. For samples with white blood cell (WBC) counts less than 36 X 10(9)/L, the +CN method was equivalent to the (postfiltration) ICSH method, whereas for WBC counts greater than 20 X 10(9)/L, the -CN method showed acceptable recovery of the mean but unacceptable imprecision. For WBC counts of 36-164 X 10(9)/L, the +CN method yielded acceptable Hb recovery with unacceptable imprecision. Hyperlipemia, resulting from addition of Intralipid directly to the blood samples, caused large errors in both H*1 methods.

Для цитирования: Malin M J, Sclafani L D, Wyatt J L. Evaluation of 24-second cyanide-containing and cyanide-free methods for whole blood hemoglobin on the Technicon H*1TM analyzer with normal and abnormal blood samples.. American journal of clinical pathology. 1989. DOI: 10.1093/ajcp/92.3.286.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.