Silicone-hydrocoated ureteral stents encrustation and biofilm formation after 3-week dwell time: results of a prospective randomized multicenter clinical study.

Barghouthy Yazeed, Wiseman Oliver, Ventimiglia Eugenio, Letendre Julien, Cloutier Jonathan, Daudon Michel, Kleinclauss Francois, Doizi Steeve, Corrales Mariela, Traxer Olivier

Сравнительное исследованиеСтатьяМногоцентровое исследованиеРандомизированное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

To explore the risk of encrustation and biofilm formation for silicone ureteral stents compared to percuflex polymer stents, through a randomized multicenter study.

PATIENTS AND

METHODS

Design, setting and participants: A Multicenter, prospective, randomized, single blind, comparative study of hydrocoated silicone stent (Coloplast Imajin® hydro) versus Percuflex™ Plus stent (Boston Scientific), in 141 patients treated by flexible URS for a kidney stone. The study had ethical committee approval in the respective hospitals. Outcome measurements and statistical analysis: Endpoints related to encrustation were biofilm formation and mineral encrustation after a period of 3-week indwelling time. They were evaluated at removal through a scoring scale of ureteral stents encrustation, infrared spectroscopy and optical microscopy of inner and outer surfaces of tips, angles and along the stent's body. Comparison was performed using ANOVA.

RESULTS

119 stents were available after removal for analysis, 56 in the silicone and 63 in the Percuflex TM Plus group. Mean dwelling duration was 21.8 days for silicone, 22.1 days for PercuflexTM Plus. There was significantly more biofilm on Percuflex™ Plus compared to silicone (1.24 ± 0.08 vs 0.93 ± 0.09, p = 0.0021), and more mineral encrustation (1.22 ± 0.10 vs 0.78 ± 0.11, p = 0.0048), respectively.

CONCLUSIONS

This multicenter randomized study shows that silicone-hydrocoated stents are less prone to encrustation than PercuflexTM Plus after a 3-week dwelling period and confirms the low encrustation potential of silicone.

Для цитирования: Barghouthy Yazeed, Wiseman Oliver, Ventimiglia Eugenio, Letendre Julien, Cloutier Jonathan, Daudon Michel, Kleinclauss Francois, Doizi Steeve, Corrales Mariela, Traxer Olivier. Silicone-hydrocoated ureteral stents encrustation and biofilm formation after 3-week dwell time: results of a prospective randomized multicenter clinical study.. World journal of urology. 2021. DOI: 10.1007/s00345-021-03646-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.