Rapid point-of-care test for diagnosis of peritonitis in peritoneal dialysis patients.
Аннотация
BACKGROUND
Periplex® is a rapid point-of-care test based on the detection of interleukin-6 (IL-6) or matrix metalloproteinase-8 (MMP-8) to diagnose peritonitis in peritoneal dialysis (PD) patients.
METHODS
This single-centre study was conducted in Singapore General Hospital from 2019 to 2022. The study recruited PD patients suspected of having peritonitis. Periplex was performed at the presentation and recovery of peritonitis. Primary outcomes were sensitivity and specificity of Periplex at presentation. The positive and negative predictive values of tests were also performed.
RESULTS
A total of 120 patients were included in the study. The mean age was 60.9 ± 14.9 years, 53% were male, 79% were Chinese and 47.5% had diabetes mellitus. Periplex was positive in all patients with peritonitis (n = 114); sensitivity of 100%; 95% confidence interval (CI): 100-100%. Periplex was falsely positive in three patients with non-infective eosinophilic peritonitis, resulting in a low specificity of 50%; 95% CI: 41.1-59.0%. Periplex had a positive predictive value of 97.4% and a negative predictive value of 100%. During recovery from peritonitis, Periplex had high specificity (93.6%) and negative predictive value (98.7%) to indicate the resolution of infection. MMP-8 was more sensitive than IL-6 in detecting peritonitis. Periplex was positive in all patients with peritonitis regardless of the types of PD solutions used.
CONCLUSIONS
Periplex had high sensitivity, and positive and negative predictive values in the diagnosis of peritonitis can be considered as a screening tool for peritonitis. Given its high specificity and negative predictive value, it may also be used to document the resolution of peritonitis.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.