Comparison between cone beam computed tomography and multislice computed tomography in diagnostic accuracy of maxillofacial fractures in dried human skull: an in vitro study.

Eskandarlou Amir, Poorolajal Jalal, Talaeipour Ahmad Reza, Talebi Sahar, Talaeipour Maziar

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

The aim of this study was to compare the accuracy of cone beam computed tomography (CBCT) and multislice computed tomography (MSCT) in diagnosing factitious fractures in maxillofacial region as well as to investigate the influence of field of view (FOV) size on the interpretation of maxillofacial fractures in CBCT.

METHODS

Simulated fractures were created in the right and left sides of five dried human skulls using a micro-saw 20 mm blade without displacement. The skulls were scanned with a spiral 16-slice MSCT scanner and a CBCT device with two FOVs (6 × 6, 15 × 15). Three observers assessed and interpreted the images judging whether fracture was present or absent. The sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and the area under the receiver operating characteristic (ROC) curve were calculated.

RESULTS

The sensitivity of CBCT 6 × 6, CBCT 15 × 15, and MSCT was 87.5%, 70.7%, and 49.7% and the specificity of three modalities was 91.7%, 88.0%, and 98.3%, respectively. The area under ROC curve was 0.90 (95% CI: 0.85, 0.94) for CBCT with FOV 6 × 6, 0.79 (95% CI: 0.75, 0.85) for CBCT with FOV 15 × 15, and 0.74 (95% CI: 0.69, 0.79) for MSCT. The Kappa value for interobserver agreement was highest for CBCT 6 × 6 among imaging modalities.

CONCLUSIONS

CBCT with small FOV had higher diagnostic accuracy in detecting simulated maxillofacial fractures than MSCT. However, additional evidence based on in vivo studies is needed to confirm the superiority of CBCT on MSCT in diagnosis of maxillofacial traumatic fractures.

Для цитирования: Eskandarlou Amir, Poorolajal Jalal, Talaeipour Ahmad Reza, Talebi Sahar, Talaeipour Maziar. Comparison between cone beam computed tomography and multislice computed tomography in diagnostic accuracy of maxillofacial fractures in dried human skull: an in vitro study.. Dental traumatology : official publication of International Association for Dental Traumatology. 2013. DOI: 10.1111/edt.12058.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.