DOI 10.1016/j.jvir.2009.12.402

Медицинская библиотека · Journal of vascular and interventional radiology : JVIR · 2010

Electromagnetic navigation for thoracic aortic stent-graft deployment: a pilot study in swine.

Abi-Jaoudeh Nadine, Glossop Neil, Dake Michael, Pritchard William F, Chiesa Alberto, Dreher Matthew R, Tang Thomas, Karanian John W, Wood Bradford J

СтатьяПеревод в очереди

Аннотация

PURPOSE: To determine the feasibility of electromagnetic tracking as a method to augment conventional imaging guidance for the safe delivery, precise positioning, and accurate deployment of thoracic aortic endografts.

MATERIALS AND

METHODS

Custom guide wires were fabricated, and the delivery catheters for thoracic aortic endoprostheses were retrofitted with integrated electromagnetic coil sensors to enable real-time endovascular tracking. Preprocedure thoracic computed tomographic (CT) angiograms were obtained after the placement of fiducial skin patches on the chest wall of three anesthetized swine, enabling automatic registration. The stent-graft deployment location target near the subclavian artery was selected on the preprocedure CT angiogram. Two steps were analyzed: advancing a tracked glidewire to the aortic arch and positioning the tracked stent-graft assembly by using electromagnetic guidance alone. Multiple CT scans were obtained to evaluate the accuracy of the electromagnetic tracking system by measuring the target registration error, which compared the actual position of the tracked devices to the displayed "virtual" electromagnetic-tracked position. Postdeployment CT angiography and necropsy helped confirm stent-graft position and subclavian artery patency.

RESULTS

A stent-graft was successfully delivered and deployed in each of the three animals by using real-time electromagnetic tracking alone. The mean fiducial registration error with autoregistration was 1.5 mm. Sixteen comparative scans were obtained to determine the target registration error, which was 4.3 mm +/- 0.97 (range, 3.0-6.0 mm) for the glidewire sensor coil. The mean target registration error for the stent-graft delivery catheter sensor coil was 2.6 mm +/- 0.7 (range, 1.9-3.8 mm). The mean deployment error for the stent-graft, defined as deployment deviation from the target, was 2.6 mm +/- 3.0.

CONCLUSIONS

Delivery and deployment of customized thoracic stent-grafts with use of electromagnetic tracking alone is feasible and accurate in swine. Combining endovascular electromagnetic tracking with conventional fluoroscopy may further improve accuracy and be a more realistic multimodality approach.

Для цитирования: Abi-Jaoudeh Nadine, Glossop Neil, Dake Michael, Pritchard William F, Chiesa Alberto, Dreher Matthew R, Tang Thomas, Karanian John W, Wood Bradford J. Electromagnetic navigation for thoracic aortic stent-graft deployment: a pilot study in swine.. Journal of vascular and interventional radiology : JVIR. 2010. DOI: 10.1016/j.jvir.2009.12.402.