DOI 10.1148/radiology.193.3.7972816

Медицинская библиотека · Radiology · 1994

Cardiac pacemaker electrodes: improved methods of extraction. Work in progress.

Smith C W, Messenger J C, Schaner S P, Renner J W

СтатьяПеревод в очереди

Аннотация

PURPOSE: To develop improved methods of extracting pacemaker electrodes from the femoral vein by using standard equipment and techniques.

MATERIALS AND

METHODS

Pacemaker electrodes were removed from 61 patients aged 21-91 years (mean, 68 years). In 25 patients (41%), traction was applied to locking stylets to remove electrodes at the subclavian vein entry site. In 30 patients, the electrode tip was freed with traction at the subclavian vein entry site. The electrode was then extracted through a sheath placed in the femoral vein. In six patients (10%), the electrode tip was freed via the femoral vein when traction was applied to grasping devices. The electrode was then removed through a femoral venous sheath.

RESULTS

Ninety-nine (95%) of 104 pacemaker electrodes were extracted completely. Two electrodes were trapped in fibrous tissue in the subclavian vein. One right atrial tear occurred.

CONCLUSION

Pacemaker electrodes can be removed effectively by using percutaneous techniques.

Для цитирования: Smith C W, Messenger J C, Schaner S P, Renner J W. Cardiac pacemaker electrodes: improved methods of extraction. Work in progress.. Radiology. 1994. DOI: 10.1148/radiology.193.3.7972816.