MedElevator Войти в исследования
← Поиск публикаций

DOI 10.2967/jnumed.121.263768

Медицинская библиотека · Journal of nuclear medicine : official publication, Society of Nuclear Medicine · 2022

68Ga-DOTATATE Prepared from Cyclotron-Produced 68Ga: An Integrated Solution from Cyclotron Vault to Safety Assessment and Diagnostic Efficacy in Neuroendocrine Cancer Patients.

Tremblay Sébastien, Beaudoin Jean-François, Bélissant Benesty Ophélie, Ait-Mohand Samia, Dumulon-Perreault Véronique, Rousseau Étienne, Turcotte Éric E, Guérin Brigitte

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

Cyclotron production of 68Ga is a promising approach to supply 68Ga radiopharmaceuticals. To validate this capability, an integrated solution for a robust synthesis of 68Ga-DOTATATE prepared from cyclotron-produced 68Ga was achieved. A retrospective comparison analysis was performed on patients who underwent PET/CT imaging after injection of DOTATATE labeled with 68Ga produced by a cyclotron or eluted from a generator to demonstrate the clinical safety and diagnostic efficacy of the radiopharmaceutical as a routine standard-of-care diagnostic tool in the clinic.

Methods

An enriched pressed 68Zn target was irradiated by a cyclotron with a proton beam set at 12.7 MeV for 100 min. The fully automated process uses an in-vault dissolution system in which a liquid distribution system transfers the dissolved target to a dedicated hot cell for the purification of 68GaCl3 and radiolabeling of DOTATATE using a cassette-based automated module. Quality control tests were performed on the resulting tracer solution. The internal radiation dose for 68Ga-DOTATATE was based on extrapolation from rat biodistribution experiments. A retrospective comparison analysis was performed on patients who underwent PET/CT imaging after injection of DOTATATE labeled with cyclotron- or generator-produced 68Ga.

Results

The synthesis of 68Ga-DOTATATE (20.7 ± 1.3 GBq) with high apparent molar activity (518 ± 32 GBq/μmol at the end of synthesis) was completed in 65 min, and the radiopharmaceutical met the requirements specified in the European Pharmacopoeia monograph on 68Ga-chloride (accelerator-produced) solution for radiolabeling. 68Ga-DOTATATE was stable for at least 5 h after formulation. The dosimetry calculated with OLINDA for cyclotron- and generator-produced 68Ga-DOTATATE was roughly equivalent. The SUVmean or SUVmax of tumoral lesions with cyclotron-produced 68Ga-DOTATATE was equivalent to that with generator-produced 68Ga. Among physiologic uptake levels, a significant difference was found in kidneys, spleen, and stomach wall, with lower values in cyclotron-produced 68Ga-DOTATATE in all cases.

Conclusion

Integrated cyclotron production achieves reliable high yields of clinical-grade 68Ga-DOTATATE. The clinical safety and imaging efficacy of cyclotron-produced 68Ga-DOTATATE in humans provide supporting evidence for its use in routine clinical practice.

Для цитирования: Tremblay Sébastien, Beaudoin Jean-François, Bélissant Benesty Ophélie, Ait-Mohand Samia, Dumulon-Perreault Véronique, Rousseau Étienne, Turcotte Éric E, Guérin Brigitte. 68Ga-DOTATATE Prepared from Cyclotron-Produced 68Ga: An Integrated Solution from Cyclotron Vault to Safety Assessment and Diagnostic Efficacy in Neuroendocrine Cancer Patients.. Journal of nuclear medicine : official publication, Society of Nuclear Medicine. 2022. DOI: 10.2967/jnumed.121.263768.