Maintaining a watertight globe while operating with a temporary keratoprosthesis.
Аннотация
PURPOSE: To demonstrate that the Landers wide-field temporary keratoprosthesis (TKP) creates a watertight seal between the tightly opposing surfaces of the TKP flange and the anterior corneal surface.
METHODS
In 25 and 20 enucleated, porcine eyes, a 7.5-mm and 8.0-mm trephination corneal hole was made, respectively, followed by open-sky capsulorhexis and lensectomy. A wide-field TKP with 7.2 mm trunk diameter was sutured securely onto each eye. Testing for a watertight seal was performed at an intraocular pressure of 100 mmHg of pre- and post-pars plana vitrectomy.
RESULTS
The Landers TKP maintained a watertight seal at an intraocular pressure of 100 mmHg in 45 of 45 eyes. After standard pars plana vitrectomy, 41 of 41 eyes maintained 100 mmHg of intraocular pressure without leakage.
CONCLUSION
The seal between the TKP and cornea is established at the horizontal interface between the TKP flange and the anterior corneal surface, not between the corneal stroma and TKP trunk. The Landers wide-field TKP is a reusable model that provides a clear posterior-segment view and effectively forms a watertight eye at a sustained intraocular pressure of 100 mmHg. This TKP may be used in ocular trauma where the corneal defect is up to 0.8 mm greater in diameter than the 7.2-mm TKP trunk.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.