Аннотация

OBJECTIVE

Approximately 25% of individualsliving with parkinsonian tremor do not respond to traditional treatments. Wearable tremor suppression devices (WTSD) provide an alternative approach, however, tremor in the fingers has not been given as much attention as tremor in the elbow and the wrist. Therefore, the objective of this study is to design a wearable tremor suppression glove that can suppress tremor simultaneously, but independently, in multiple hand joints without restricting the user's voluntary motion.

METHODS

A WTSD was designed for managing tremor in the index finger metacarpophalangeal (MCP) joint, thumb MCP joint, and the wrist. The prototype was tested and assessed on a participant living with parkinsonian tremor.

RESULTS

The experimental evaluation showed an overall suppression of 73.1%, 80.7%, and 85.5% in resting tremor, 70.2%, 79.5%, and 81% in postural tremor, and 60.0%, 58.7%, and 65.0% in kinetic tremor in the index finger MCP joint, the thumb MCP joint, and the wrist, respectively.

CONCLUSION

This first assessment of a WTSD for people living with Parkinson's disease provides confirmation of the feasibility of the approach. The next step requires a comprehensive validation on a broader population in order to evaluate the performance of the WTSD.

SIGNIFICANCE: This study demonstrates the feasibility of using a WTSD to manage hand and finger tremor. The device enriches the field of upper-limb tremor management, as the first WTSD for multiple joints of the hand.

Для цитирования: Zhou Yue, Ibrahim Anas, Hardy Kenneth G, Jenkins Mary E, Naish Michael D, Trejos Ana Luisa. Design and Preliminary Performance Assessment of a Wearable Tremor Suppression Glove.. IEEE transactions on bio-medical engineering. 2021. DOI: 10.1109/tbme.2021.3080622.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.