Randomized Trial of Assisted Hybrid Closed-Loop Therapy Versus Sensor-Augmented Pump Therapy in Pregnancy.
Аннотация
Objective
Examine gestational safety, glycemic and health outcomes, of a hybrid closed-loop (HCL) system without pregnancy-specific glucose targets. Research
Design
This was a pilot feasibility investigator-initiated, two-site, single-blind, randomized controlled trial of sensor-augmented pump therapy (SAPT) versus HCL therapy in type 1 diabetes pregnancies. Participants were enrolled in the first trimester and randomized at 14-18 weeks of gestation and used SAPT or HCL until 4-6 weeks postpartum. We compared continuous glucose monitoring (CGM) metrics, severe hypoglycemia (SH), diabetic ketoacidosis (DKA), adverse skin reactions, and pregnancy outcomes between groups.
Results
Baseline characteristics were similar between groups (n = 11 HCL and n = 12 SAPT). There was no SH or DKA episode after randomization. Time spent <54 mg/dL did not differ between groups. Time spent <63 mg/dL decreased in both groups, significantly in the HCL group (3.5% [1.3% standard error] second trimester and 2.8% [1.3%] third trimester vs. 7.9% [1.3%] run-in phase, P < 0.05 for both). Mean sensor glucose was lower with SAPT compared to HCL therapy in the third trimester (119 [4] mg/dL SAPT vs. 132 [4] mg/dL HCL, P < 0.05). Third trimester time-in-range (TIR; 63-140 mg/dL) increased with SAPT (68.2% [3.1%] vs. 64.3% [3.1%] run-in phase, P < 0.05). Gestational health outcomes did not differ between groups. The HCL group used assistive techniques, such as fake carbohydrate boluses and exiting HCL overnight.
Conclusions
CGM within group differences were seen for time <63 mg/dL favoring HCL therapy and TIR favoring SAPT (third trimester vs. baseline). Safety and adverse pregnancy outcomes were similar between groups.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.