Concomitant Intra-Aortic Balloon Pumping Significantly Reduces Left Ventricular Pressure during Central Veno-Arterial Extracorporeal Membrane Oxygenation-Results from a Large Animal Model.
Аннотация
(1)
Introduction
Simultaneous ECMO and IABP therapy is frequently used. Haemodynamic changes responsible for the success of the concomitant mechanical circulatory support system approach are rarely investigated. In a large-animal model, we analysed haemodynamic parameters before and during ECMO therapy, comparing central and peripheral ECMO circulation with and without simultaneous IABP support. (2)
Methods
Thirty-three female pigs were divided into five groups: (1) SHAM, (2) (peripheral)ECMO(-)IABP, (3) (p)ECMO(+)IABP, (4) (central)ECMO(-)IABP, and (5) (c)ECMO(+)IABP. Pigs were cannulated in accordance with the group and supported with ECMO (±IABP) for 10 h. Systemic haemodynamics, cardiac index (CI), and coronary and carotid artery blood flow were determined before, directly after, and at five and ten hours on extracorporeal support. Systemic inflammation (IL-6; IL-10; TNFα; IFNγ), immune response (NETs; cf-DNA), and endothelial injury (ET-1) were also measured. (3)
Results
IABP support during antegrade ECMO circulation led to a significant reduction of left ventricular pressure in comparison to retrograde flow in (p)ECMO(-)IABP and (p)ECMO(+)IABP. Blood flow in the left anterior coronary and carotid artery was not affected by extracorporeal circulation. (4)
Conclusions
Concomitant central ECMO and IABP therapy leads to significant reduction of intracavitary cardiac pressure, reduces cardiac work, and might therefore contribute to improved recovery in ECMO patients.